ANIMAS VIBE
Report
- Report Number
- 2531779-2013-05469
- Event Type
- Malfunction
- Date Received
- April 29, 2013
- Date of Event
- March 18, 2013
- Report Date
- April 5, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(6). THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 (B)(4) - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS CONFIRMED TO BE TORN BETWEEN THE UP ARROW AND CONTRAST KEYPAD BUTTONS. EVALUATION REVEALED THAT THE KEYPAD BUTTONS WERE RESPONSIVE TO USER INPUT. THERE WAS NO EVIDENCE OF CONTAMINATION FOUND UNDER THE KEY CONTACTS. UNRELATED TO THE COMPLAINT, THE BOLUS BUTTON WAS FOUND TO BE DETACHED FROM THE PUMP, EXPOSING THE KEY CONTACTS AND THEREFORE, THE AUDIBLE ALARM READING WAS UNABLE TO BE OBTAINED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE DISTRIBUTOR CONTACTED ANIMAS ON (B)(6) 2013, ALLEGING THE KEYPAD WAS PEELED/TORN/WORN. NO FURTHER INFORMATION WAS AVAILABLE. A DAMAGED KEYPAD RUBBER WILL PERMIT CONTAMINATION TO PERMEATE THE BUTTONS WHICH WILL HAVE A NEGATIVE IMPACT ON BUTTON FUNCTION. THIS SITUATION IS NOT LIKELY TO RESULT IN AN ADVERSE EVENT AS THE DAMAGED KEYPAD RUBBER SHOULD BE CLEARLY VISIBLE AND PROHIBIT THE USE OF THE KEYPAD BUTTONS. USERS MAY EXPECT KEYPAD DAMAGE DURING NORMAL WEAR AND THE OWNER'S BOOKLET INSTRUCTS THE USER TO CONTACT CUSTOMER SERVICE IF THE USER SUSPECTS THE PUMP MAY BE DAMAGED. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED KEYPAD ISSUE REMAINED UNRESOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 183420 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |