PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2013-92912
- Event Type
- Malfunction
- Date Received
- April 29, 2013
- Date of Event
- April 7, 2013
- Report Date
- April 7, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP RECEIVED WITH PRIME ALARMS DURING PRIME TEST AND PRIME ALARMS DURING BASIC OCCLUSION TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. UNABLE TO PERFORM OCCLUSION AND EXCESSIVE NO DELIVERY ALARM TEST DUE TO PRIME ALARMS. THE INSULIN PUMP PASSED DISPLACEMENT TEST. MOTOR WAS INSPECTED AND TESTED. NO MOTOR ANOMALY NOTED. CRACKED RESERVOIR TUBE LIP, CRACKED CASE NEAR DISPLAY WINDOW CORNERS AND CRACKED BATTERY TUBE THREADS NOTED DURING VISUAL INSPECTION.
CUSTOMER REPORTED THAT THE INSULIN PUMP DRIVE SUPPORT CAP IS STICKING OUT. NOTHING FURTHER WAS REPORTED.