FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3081828 · Received April 29, 2013

Report

Report Number
3004209178-2013-92912
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
April 7, 2013
Report Date
April 7, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP RECEIVED WITH PRIME ALARMS DURING PRIME TEST AND PRIME ALARMS DURING BASIC OCCLUSION TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. UNABLE TO PERFORM OCCLUSION AND EXCESSIVE NO DELIVERY ALARM TEST DUE TO PRIME ALARMS. THE INSULIN PUMP PASSED DISPLACEMENT TEST. MOTOR WAS INSPECTED AND TESTED. NO MOTOR ANOMALY NOTED. CRACKED RESERVOIR TUBE LIP, CRACKED CASE NEAR DISPLAY WINDOW CORNERS AND CRACKED BATTERY TUBE THREADS NOTED DURING VISUAL INSPECTION.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP DRIVE SUPPORT CAP IS STICKING OUT. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
185296 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAL

Patients

Seq Age Sex Outcome Treatment
1 37 YR