FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3081811 · Received April 29, 2013

Report

Report Number
3004209178-2013-92904
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
April 6, 2013
Report Date
April 6, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. UNABLE TO PERFORM THE OCCLUSION, EXCESSIVE NO DELIVERY ALARM TESTS DUE TO PRIME FILL ANOMALY. UNIT HAD MISSING END CAP STICKER. HOWEVER, UNIT PASSED THE DISPLACEMENT TEST.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP DRIVE SUPPORT CAP IS STICKING OUT. NOTHING FURTHER WAS REPORTED.