FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 3081758 · Received April 29, 2013

Report

Report Number
2531779-2013-05432
Event Type
Malfunction
Date Received
April 29, 2013
Report Date
April 3, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW UP #1 SUBMITTED: (B)(4) 2013. DEVICE EVALUATION: ON INVESTIGATION, THE PUMP HAD AUDIBLE TONES AND VIBRATORY FUNCTION WHEN POWERED ON; HOWEVER, THE DISPLAY SCREEN WAS BLANK. AN ALARM WAS UNABLE TO BE CLEARED. INVESTIGATION CONFIRMED A COMPONENT FAILURE ON THE PRINTED CIRCUIT BOARD. COMPLETE TESTING WAS NOT ABLE TO BE COMPLETED DUE TO THE BLANK DISPLAY.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(4).

Description of Event or Problem · 1

THE DISTRIBUTOR CONTACTED ANIMAS ON (B)(6) 2013 AND REPORTED THE UP, DOWN, AND OK KEYPAD BUTTONS WERE UNRESPONSIVE. THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGED KEYPAD RESPONSE ISSUE.