COBAS INTEGRA 800
Report
- Report Number
- 1823260-2013-02588
- Event Type
- Malfunction
- Date Received
- April 29, 2013
- Date of Event
- April 6, 2013
- Report Date
- April 26, 2013
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JGS
- PMA / PMN Number
- K951595
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA.
THE CUSTOMER REPORTED THAT THEY RECEIVED COMPLAINTS FOR LOW ION SELECTIVE (ISE) SODIUM RESULTS ON PATIENT SAMPLES. THE CUSTOMER QUESTIONED A TOTAL OF TEN SAMPLES TESTED FOR ISE SODIUM. QUALITY CONTROLS AND CALIBRATION WERE ACCEPTABLE PRIOR TO RUNNING THE PATIENT SAMPLES, BUT WHEN QUALITY CONTROLS WERE CHECKED AFTER THE EVENT, THEY WERE OUT OF RANGE LOW. CALIBRATIONS WERE ALSO PERFORMED SEVERAL TIMES AFTER THE EVENT WITH THE LAST CALIBRATION FAILING OUT OF RANGE. THE CUSTOMER TRIED TO PERFORM ELECTRODE ACTIVATION, BUT THIS DID NOT RESOLVE THE ISSUE. THE CUSTOMER PROVIDED DATA FOR THREE PATIENT SAMPLES AND TWO WERE FOUND TO HAVE ERRONEOUS RESULTS. ALL SAMPLES WERE REPEATED ON INTEGRA 400 PLUS SERIAL NUMBER (B)(4) AND THE REPEAT RESULTS WERE BELIEVED TO BE CORRECT. SAMPLE ONE INITIALLY RESULTED AS 129 MMOL/L AND THIS VALUE WAS RELEASED OUTSIDE OF THE LABORATORY. THE SAMPLE WAS REPEATED AND RESULTED AS 137 MMOL/L. SAMPLE TWO INITIALLY RESULTED AS 129 MMOL/L AND THIS VALUE WAS RELEASED OUTSIDE OF THE LABORATORY. THE SAMPLE WAS REPEATED AND RESULTED AS 137 MMOL/L. THE PATIENTS WERE NOT ADVERSELY AFFECTED BY THE EVENT. THE ISE SODIUM ELECTRODE LOT NUMBER AND EXPIRATION DATE WERE NOT PROVIDED. THE FIELD SERVICE REPRESENTATIVE INDICATED THAT THE ISSUE WAS CAUSED BY A DAMAGED ELECTRODE. THE CAUSE OF THE ELECTRODE FAILURE COULD NOT BE DETERMINED. THE FIELD SERVICE REPRESENTATIVE CLARIFIED THAT THERE WAS NO PHYSICAL DAMAGE TO THE ELECTRODE. HE REPLACED THE SODIUM ELECTRODE. THE CUSTOMER CALIBRATED, RAN QUALITY CONTROLS, AND PRECISION; ALL WERE OK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 184499 | COBAS INTEGRA 800 | CLINICAL CHEMISTRY ANALYZER | JGS | ROCHE DIAGNOSTICS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |