FDA Adverse Event Malfunction Summary report: N

RESERVOIR 1.8ML

MDR report key: 3081700 · Received April 29, 2013

Report

Report Number
3004209178-2013-92828
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
April 6, 2013
Report Date
April 6, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K001828
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT SHE HAS BEEN HAVING UNEXPLAINED HIGH BLOOD GLUCOSE. CUSTOMER STATED THAT HER RESERVOIR LEAKED INSULIN PAST BOTH O-RINGS DURING NORMAL USE INTO THE INSULIN PUMP RESERVOIR COMPARTMENT. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
184466 RESERVOIR 1.8ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-326A

Patients

Seq Age Sex Outcome Treatment
1 46 YR