FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3081685 · Received April 29, 2013

Report

Report Number
3004209178-2013-92849
Event Type
Injury
Date Received
April 29, 2013
Date of Event
April 1, 2013
Report Date
April 5, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT SHE HAD UNEXPLAINED HIGH BLOOD GLUCOSE. CUSTOMER WENT TO THE EMERGENCY ROOM AND WAS HOSPITALIZED. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE EMERGENCY ROOM VISIT WAS 502 MG/DL. CUSTOMER STATED THAT HER INSULIN PUMP DELIVERED INSULIN WILE SHE WAS SLEEPING, AND CAUSE A LOW BLOOD GLUCOSE OF 40 MG/DL. CUSTOMER ALSO STATED THAT THE DRIVE SUPPORT CAP BROKE OFF ON HER INSULIN PUMP. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
185183 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 53 YR Hospitalization