FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3081681 · Received April 29, 2013

Report

Report Number
3004209178-2013-92777
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
April 23, 2013
Report Date
April 23, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED DISPLACEMENT TEST, BASIC OCCLUSION AND PRIME TESTS. THE INSULIN PUMP RECEIVED STUCK IN MOTOR ERROR ALARM LOOP DURING BOLUS DELIVERY. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE MOTOR MAY HAVE HAD AN INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING OUR TESTING.

Description of Event or Problem · 1

IT IS REPORTED THAT INSULIN PUMP ALARMED MOTOR ERROR. NO EXPOSURE TO HIGH MAGNETIC FIELDS. THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. MOTOR ERROR OCCURED DURING PRIME. CUSTOMER CHANGE INFUSION SET AND RESERVOIR, MOTOR ERROR ALARM PERSISTED. ASK CUSTOMER TO RETURN THE INSULIN PUMP FOR ANALYSIS. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
185628 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-522WWB

Patients

Seq Age Sex Outcome Treatment
1