ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-05414
- Event Type
- Malfunction
- Date Received
- April 26, 2013
- Date of Event
- March 18, 2013
- Report Date
- April 3, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1 DATE OF SUBMISSION 08/01/2013 DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT. EVALUATION REVEALED THAT THE KEYPAD BUTTONS WERE RESPONSIVE TO USER INPUT. THERE WAS NO EVIDENCE OF CONTAMINATION FOUND UNDER THE KEY CONTACTS. THE REPORTED ISSUE WAS NOT DUPLICATED DURING TESTING. UNRELATED TO THE COMPLAINT, THE PINS ON THE VIBRATORY MOTOR WERE NOT MAKING AN ELECTRICAL CONNECTION ON THE PCB BOARD. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
ON (B)(6) 2013, THE DISTRIBUTOR CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. REPORTEDLY, THE UP, DOWN, AND OK BUTTONS WERE UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.