FDA Adverse Event Malfunction Summary report: N

8 MM MCS TIP COVER ACCESSORY

MDR report key: 3081313 · Received April 26, 2013

Report

Report Number
2955842-2013-01419
Event Type
Malfunction
Date Received
April 26, 2013
Date of Event
March 27, 2013
Report Date
April 1, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TIP COVER HAS NOT BEEN RETURNED FOR EVALUATION, THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE ACCESSORY IS RETURNED (POST ENGINEERING EVALUATION) OR IF ADDITIONAL INFORMATION IS RECEIVED. ON (B)(4) 2013, INTUITIVE SURGICAL INC. (ISI) CONTACTED THE DISTRIBUTOR TO OBTAIN ADDITIONAL INFORMATION REGARDING THE REPORTED EVENT. THE MCS TIP COVER ACCESSORY WAS NOT AVAILABLE FOR RETURN TO ISI FOR EVALUATION. THE DISTRIBUTOR INDICATED THAT THE MCS TIP COVER ACCESSORY WAS INSPECTED PRIOR TO USE AND WAS INSTALLED ON THE MCS INSTRUMENT USING THE INSTALLATION TOOL. THE DISTRIBUTOR WAS UNABLE TO INDICATE IF ELECTROLUBE WAS APPLIED TO THE MCS INSTRUMENT PRIOR TO INSTALLATION OF THE MCS TIP COVER ACCESSORY. THE CANNULA WAS INSPECTED WITH A PIN GAGE PRIOR TO USE. THE DISTRIBUTOR WAS UNABLE TO PROVIDE THE ELECTROSURGICAL GENERATOR UNIT (ESU) SETTINGS USED DURING THE REPORTED SURGICAL PROCEDURE. THE DISTRIBUTOR INDICATED THAT THE MCS INSTRUMENT COLLIDED WITH ANOTHER INSTRUMENT DURING THE SURGICAL PROCEDURE. HOWEVER, THE DISTRIBUTOR WAS UNABLE TO PROVIDE DETAILS OF THE DAMAGE FOUND ON THE MCS TIP COVER ACCESSORY. THE DISTRIBUTOR INDICATED THAT THE SURGICAL PROCEDURE WAS RECORDED. HOWEVER, THE DISTRIBUTOR DECLINED TO PROVIDE A COPY OF THE VIDEO TO ISI FOR EVALUATION. THE DISTRIBUTOR DENIED THAT THE PATIENT SUSTAINED ANY INJURIES OR COMPLICATIONS BECAUSE OF THE REPORTED MCS TIP COVER ACCESSORY ISSUE. THE DISTRIBUTOR INDICATED THAT THE PATIENT WAS DISCHARGED FROM THE HOSPITAL WITH NO REPORTED PROBLEMS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S HYSTERECTOMY PROCEDURE, THE SURGEON OBSERVED A SPARK COMING FROM THE MONOPOLAR CURVED SCISSORS (MCS) TIP COVER ACCESSORY INSTALLED ON THE MCS INSTRUMENT. EXAMINATION OF THE MCS TIP COVER ACCESSORY BY THE SURGICAL STAFF FOUND AN UNSPECIFIED BROKEN PORTION. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED WITH A REPLACEMENT MCS TIP COVER ACCESSORY AND WITHOUT ANY OTHER PROBLEMS. NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
182587 8 MM MCS TIP COVER ACCESSORY ENDOSCOPIC ELECTROSURGICAL INSTRUMENT ACCESSORY NAY INTUITIVE SURGICAL,INC. 400180-10

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SURG SYS INST., ACCESSORIES, & ESU