Report Number
2531779-2013-05401
Event Type
Malfunction
Date Received
April 26, 2013
Report Date
April 9, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 2 DATE OF SUBMISSION (B)(4) 2013: CORRECTION: INVESTIGATION CODES WERE INADVERTENTLY OMITTED FOR THE ORIGINAL INVESTIGATION RESULTS. UPDATED SUPPLEMENTAL TO ADD EVALUATION CODES.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS BUT EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION (B)(4) 2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE DISPLAY SCREEN WAS FOUND TO BE DIM AND DISCOLORED. A TEST SCREEN WAS INSERTED FOR TESTING AND WAS FOUND TO FUNCTION PROPERLY. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A CRACKED BATTERY COMPARTMENT.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY SCREEN ISSUE. IT WAS REPORTED THAT THE DISPLAY WAS FADED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE DEVICE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
180910 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 63 YR