I-STAT TROPONIN (CTNI) CARTRIDGE
Report
- Report Number
- 2245578-2013-00068
- Event Type
- Malfunction
- Date Received
- April 26, 2013
- Date of Event
- April 5, 2013
- Report Date
- April 11, 2013
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- MMI
- PMA / PMN Number
- K031739
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). THE INVESTIGATION WAS COMPLETED ON (B)(4) 2013. RETAIN AND RETURNED CARTRIDGES WERE TESTED AND ARE FUNCTIONING ACCORDING TO SPECIFICATION.
(B)(4).
NA
ON (B)(6) 2013, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT RESULT ON (B)(6) FEMALE PRESENTED IN ER WITH CHEST PAINS/SHORTNESS OF BREATH. TEST DATE: (B)(6) 2013METHOD TIME RESULT I-STAT 10:24AM 0.11 (POSITIVE RESULT)I-STAT 10:39AM 0.00 (NEGATIVE RESULT)VISTA 11:13AM <0.015 (NEGATIVE RESULT)THERE WAS NO OTHER PATIENT INFORMATION RECIEVED AT THIS TIME. BASED ON THE INFORMATION AVAILABLE AT THE TIME, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 182056 | I-STAT TROPONIN (CTNI) CARTRIDGE | CTNI CARTRIDGE | MMI | ABBOTT POINT OF CARE | NA | U13033A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR |