I-STAT TROPONIN (CTNI) CARTRIDGE

MDR report key: 3081135 · Received April 26, 2013

Report

Report Number
2245578-2013-00067
Event Type
Malfunction
Date Received
April 26, 2013
Date of Event
April 6, 2013
Report Date
April 11, 2013
Manufacturer
ABBOTT POINT OF CARE
Product Code
MMI
PMA / PMN Number
K031739
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE INVESTIGATION WAS COMPLETED ON (B)(4) 2013. RETAIN AND RETURNED CARTRIDGES WERE TESTED AND ARE FUNCTIONING ACCORDING TO SPECIFICATION.

Description of Event or Problem · 1

NA

Description of Event or Problem · 1

ON (B)(6) 2013, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT RESULT ON A (B)(6) MALE PRESENTED IN ER WITH CHEST PAINS/SHORTNESS OF BREATH. TEST DATE: (B)(6) 2013: METHOD: TIME: RESULT: I-STAT, 12:35AM, 0.94 (POSITIVE RESULT). I-STAT, 12:58AM, 0.00 (NEGATIVE RESULT). VISTA, 01:08AM, <0.015 (NEGATIVE RESULT). EKG SHOWED PREMATURE VENTRICULAR COMPLEX. THERE WAS NO OTHER PATIENT INFORMATION RECEIVED AT THIS TIME. BASED ON THE INFORMATION AVAILABLE AT THE TIME, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
183069 I-STAT TROPONIN (CTNI) CARTRIDGE CTNI CARTRIDGE MMI ABBOTT POINT OF CARE NA U13033A

Patients

Seq Age Sex Outcome Treatment
1 62 YR