FDA Adverse Event
Malfunction
Summary report: N
IN TOUCH ZU
MDR report key: 3081044
·
Received April 26, 2013
Report
- Report Number
- 0001831750-2013-03780
- Event Type
- Malfunction
- Date Received
- April 26, 2013
- Date of Event
- April 5, 2013
- Report Date
- April 5, 2013
- Manufacturer
- STRYKER MEDICAL-KALAMAZOO
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
IT WAS REPORTED VIA REPAIR WORK ORDER THAT THE FOOTBOARD HAD INTERMITTENT POWER AS THE FOOTBOARD CONNECTOR PINS WERE OUT OF ALIGNMENT. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 182956 | IN TOUCH ZU | BED, AC-POWERED ADJUSTABLE HOSPITAL | FNL | STRYKER MEDICAL-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |