FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3081027 · Received April 26, 2013

Report

Report Number
1416980-2013-10525
Event Type
Malfunction
Date Received
April 26, 2013
Date of Event
March 15, 2013
Report Date
April 5, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS ONE OR MORE CYCLES ADVANCED TO THE NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING THE EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED DURING THE THERAPY INITIATED ON (B)(6) 2013 AT 12:09:22. DURING NIGHT DRAIN CYCLE FOUR, THE PATIENT'S ULTRAFILTRATION READING WAS 1871ML, INDICATING THE HOME PATIENT (HP) DRAINED 1871ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2500ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
182477 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 74 YR