ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2013-01996
- Event Type
- Malfunction
- Date Received
- April 26, 2013
- Date of Event
- March 28, 2013
- Report Date
- April 9, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ASKED FOR BUT UNKNOWN OR NOT PROVIDED DURING INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
(B)(4). THE DEVICE WAS RETURNED WITH THE TISSUE PAD MELTED AND PARTIALLY DETACHED, NOT COMPLETED GONE AS REPORTED. THE DEVICE WAS CONNECTED TO A TEST HANDPIECE AND GENERATOR AND IT DID ACTIVATE DURING FUNCTIONAL TESTING. PROBABLE CAUSES OF TISSUE PAD DAMAGE ARE APPLYING PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. AVOID ACTIVATING THE BLADE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO PREVENT DAMAGE TO THE TISSUE PAD.
IT WAS REPORTED THAT DURING A LAPAROSCOPY PROCEDURE, THE PAD MELTED OFF AND WAS RETRIEVED. THERE WERE NO PATIENT CONSEQUENCES. CASE COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE. ONE DEVICE WILL BE RETURNED.