FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3080827 · Received April 12, 2013

Report

Report Number
3008642652-2013-00994
Event Type
Malfunction
Date Received
April 12, 2013
Date of Event
March 6, 2013
Report Date
April 5, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (SQUEALING / WILL NOT POWER ON) WAS CONFIRMED. AS RECEIVED, THE MONITOR WOULD NOT POWER ON. THE CAUSE OF THE INABILITY TO POWER ON WAS A FLASH MEMORY FAILURE (COMPONENTS U102 AND U105) ON THE C/A BOARD SN (B)(4). THE FLASH MEMORY HAD AN INTERMITTENT BGA CONNECTION, WHICH WAS DISCOVERED THROUGH THE USE OF A TARGET (OR FLASH) MEMORY TEST. THE INTERMITTENT CONNECTION IS LIKELY DUE TO A FRACTURED SOLDER CONNECTION INDUCED BY MECHANICAL STRESS. THE EXACT SOURCE OF THE MECHANICAL STRESS HAS NOT BEEN POSITIVELY IDENTIFIED BUT THE FRACTURED CONNECTION MAY HAVE BEEN ACCELERATED THROUGH ROUGH HANDLING. ROOT CAUSE ANALYSIS OF SIMILAR EVENTS FOUND THAT EXCESSIVE STRAIN ON THE PRINTED CIRCUIT ASSEMBLY (PCA) FROM SEVERE IMPACT CAN CAUSE FRACTURED SOLDER CONNECTIONS. (B)(4). NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE MONITOR. THE PT RECEIVED A REPLACEMENT MONITOR.

Description of Event or Problem · 1

A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE MONITOR EMITTED A HIGH PITCH SQUEAL AND WAS UNRESPONSIVE. THE PT WAS ISSUED A REPLACEMENT MONITOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
157864 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 80 YR