FDA Adverse Event Injury Summary report: N

ICL (IMPLANTABLE COLLAMER LENS)

MDR report key: 3080807 · Received April 26, 2013

Report

Report Number
2023826-2013-00317
Event Type
Injury
Date Received
April 26, 2013
Date of Event
November 10, 2012
Report Date
April 5, 2013
Manufacturer
STAAR SURGICAL COMPANY
Product Code
MTA
PMA / PMN Number
P030016
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4) - NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT; SIZE INCORRECT FOR PATIENT. (B)(4).

Description of Event or Problem · 1

THE REPORTER INDICATED THE SURGEON IMPLANTED A 11.5MM ICM115V4 IMPLANTABLE COLLAMER LENS IN THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2012. THE LENS WAS EXPLANTED ON (B)(6) 2012 DUE TO LOW VAULTING. THE ICL WAS EXCHANGED FOR A LONGER LENS WHICH RESOLVED THE PROBLEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
182306 ICL (IMPLANTABLE COLLAMER LENS) INTRAOCULAR LENS MTA STAAR SURGICAL COMPANY ICM115V4 NA

Patients

Seq Age Sex Outcome Treatment
1 30 YR Required Intervention INJECTOR MODEL AND LOT NUMBER UNK| CARTRIDGE MODEL AND LOT NUMBER UNK