LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2013-00906
- Event Type
- Malfunction
- Date Received
- April 12, 2013
- Date of Event
- March 16, 2013
- Report Date
- April 1, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGE CABLE) HAS BEEN CONFIRMED. UPON EVALUATION, THE CABLE CONNECTING THE REAR THERAPY ELECTRODE (TE) AND THE DISTRIBUTION NODE (DN) WAS PULLED FROM THE DN STRAIN RELIEF, DAMAGING INTERNAL WIRES. THE ROOT CAUSE OF THE DAMAGED CABLE AND WIRES CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CABLE AND WIRES. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
A PATIENT SERVICE REPRESENTATIVE (PSR) CONTACTED ZOLL CUSTOMER SUPPORT WHILE ASSISTING A (B)(6) MALE PATIENT TO REPORT A DAMAGED CABLE ON THE PATIENT'S ELECTRODE BELT. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 159636 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 27 YR |