FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3080796 · Received April 12, 2013

Report

Report Number
3008642652-2013-00984
Event Type
Malfunction
Date Received
April 12, 2013
Date of Event
March 23, 2013
Report Date
April 5, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WON'T POWER ON) HAS BEEN CONFIRMED. AS RECEIVED, THE CHARGER/ MODEM WOULD NOT POWER ON. UPON EVALUATION, THE POWER SUPPLY UNIT WAS DEFECTIVE. THE CAUSE OF THE INABILITY TO POWER ON IS THE DEFECTIVE POWER SUPPLY UNIT. THE ROOT CAUSE OF THE DEFECTIVE POWER SUPPLY UNIT CANNOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE POWER SUPPLY. THE PATIENT WAS ISSUED A REPLACEMENT BATTERY CHARGER.

Description of Event or Problem · 1

A PATIENT SERVICE REPRESENTATIVE (PSR) CONTACTED ZOLL CUSTOMER SUPPORT WHILE ASSISTING A (B)(6) MALE PATIENT TO REPORT THAT THE BATTERY CHARGER WOLD NOT POWER ON. THE PATIENT WAS ISSUED A REPLACEMENT CHARGER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
159635 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 85 YR