FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3080790 · Received April 11, 2013

Report

Report Number
3008642652-2013-01036
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
March 11, 2013
Report Date
April 10, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. UPON EVAL, THERE WAS AN OPEN PULSE WIRE INSIDE IN THE CABLE CONNECTING THE DISTRIBUTION NODE (DN) TO ECG ELECTRODE B. THE CAUSE FOR THE TEST FAILURE IS THE OPEN PULSE WIRE. THE ROOT CAUSE OF THE OPEN WIRE CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE OPEN WIRE. THE LAST PT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.

Description of Event or Problem · 1

REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, BELT SN (B)(4) FAILED THE INCOMING TE RECOGNITION TEST. THE LAST PT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
157218 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA