Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. UPON EVAL, THERE WAS AN OPEN PULSE WIRE INSIDE IN THE CABLE CONNECTING THE DISTRIBUTION NODE (DN) TO ECG ELECTRODE B. THE CAUSE FOR THE TEST FAILURE IS THE OPEN PULSE WIRE. THE ROOT CAUSE OF THE OPEN WIRE CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE OPEN WIRE. THE LAST PT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.
Description of Event or Problem · 1
REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, BELT SN (B)(4) FAILED THE INCOMING TE RECOGNITION TEST. THE LAST PT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.