ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-05379
- Event Type
- Malfunction
- Date Received
- April 26, 2013
- Report Date
- April 10, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1 06/10/2013 ¿ DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THERE WAS NO VISIBLE DAMAGE TO THE KEYPAD. THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND CONTAMINATION WAS OBSERVED UNDER THE BUTTON CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. IT WAS REPORTED THAT THE UP, DOWN, AND OK BUTTON WERE INTERMITTENTLY UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 180820 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |