FDA Adverse Event Summary report: N

BRILLIANCE 16 AIR

MDR report key: 3080514 · Received April 24, 2013

Report

Report Number
1525965-2013-00091
Date Received
April 24, 2013
Report Date
March 29, 2013
Manufacturer
PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC.
Product Code
JAK
PMA / PMN Number
K012009
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NOTE: WE HAVE NOT COMPLETED OUR INVESTIGATION OF THIS EVENT. WE WILL FILE A F/U MDR AT THE COMPLETION OF THE INVESTIGATION. (B)(4). INITIAL MDR MAILED ON (B)(4) 2013.

Description of Event or Problem · 1

PHILIPS RECEIVED A REPORT FROM A CUSTOMER REPORTING THAT THEY WERE PERFORMING A SURVIEW SCAN ON A PT AND REPORTED THAT THE PT SUPPORT TOP LOWERED UNEXPECTEDLY UNTIL IT CONTACTED THE GANTRY FRONT COVER AND THE SYSTEM E-STOP OPENED, WHICH HALTED THE PROCEDURE AND TERMINATED THE EXPOSURE. NO ONE WAS INJURED BY THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
177594 BRILLIANCE 16 AIR JAK PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC. 728246

Patients

Seq Age Sex Outcome Treatment
1