FDA Adverse Event Injury Summary report: N

UNIFY CRT-D

MDR report key: 3080212 · Received April 26, 2013

Report

Report Number
2938836-2013-01011
Event Type
Injury
Date Received
April 26, 2013
Date of Event
March 15, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NIK
PMA / PMN Number
P030054
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT INAPPROPRIATE ATP THERAPY WAS OBSERVED DUE TO SVT. THE DEVICE WAS REPROGRAMMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
182362 UNIFY CRT-D IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION CD3231-40 NA

Patients

Seq Age Sex Outcome Treatment
1 74 YR Required Intervention