FDA Adverse Event Injury Summary report: N

RIATA PASSIVE FIXATION

MDR report key: 3080207 · Received April 26, 2013

Report

Report Number
2938836-2013-00895
Event Type
Injury
Date Received
April 26, 2013
Date of Event
March 8, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A PARTIAL LEAD WITH THE CONNECTOR PIN MEASURING 16.2CM WAS RETURNED FOR ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 11.2-11.4CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS INTACT AT THIS LOCATION.

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD WAS CAPPED DUE TO OVERSENSING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
182849 RIATA PASSIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1570/65 NA

Patients

Seq Age Sex Outcome Treatment
1 63 YR Required Intervention