FDA Adverse Event Injury Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 3080175 · Received April 26, 2013

Report

Report Number
2938836-2013-00943
Event Type
Injury
Date Received
April 26, 2013
Date of Event
February 14, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
P950022
Removal / Correction Number
Z0458
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A PARTIAL LEAD WAS RETURNED FOR ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 7.5-8.0CM AND 43.9-44.9CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. THE ETFE COATING WAS INTACT AT THESE LOCATIONS.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Description of Event or Problem · 1

IT WAS REPORTED THAT EXTERNALIZED CONDUCTORS WERE OBSERVED. NO ELECTRICAL ANOMALIES WERE NOTED. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
182218 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7000/65 NA

Patients

Seq Age Sex Outcome Treatment
1 62 YR Required Intervention