FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 3080115 · Received April 26, 2013

Report

Report Number
2938836-2013-00958
Event Type
Injury
Date Received
April 26, 2013
Date of Event
February 27, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED EXTERNAL INSULATION ABRASION WAS FOUND AT 19.0-20.5CM FROM THE CONNECTOR PIN CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS ABRADED AND THE RE CONDUCTOR WAS FRACTURED AT THIS LOCATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE SHOCKS. UPON EXPLANT, AN INSULATION ABRASION CONSISTENT WITH FRICTION TO CAN WAS OBSERVED. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
180888 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120/65 NA

Patients

Seq Age Sex Outcome Treatment
1 33 YR Required Intervention (B)(4)