FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3080111 · Received April 26, 2013

Report

Report Number
3004209178-2013-92803
Event Type
Injury
Date Received
April 26, 2013
Date of Event
April 1, 2013
Report Date
April 5, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP UNABLE TO PRIME DURING PRIME TEST. THE INSULIN PUMP RECEIVED WITH MOTOR ERROR ALARMS DURING BASIC OCCLUSION TEST DUE TO FAULTY FORCE SENSOR. UNABLE TO PERFORM OCCLUSION TEST AND EXCESSIVE NO DELIVERY TEST DUE TO PRIME/FILL ANOMALY. HOWEVER, THE INSULIN PUMP PASSED THE DISPLACEMENT TEST.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO HIGH BLOOD SUGARS AND DKA. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF HOSPITALIZATION WAS OVER 6000 MG/DL. CALLER STATED THAT THE CUSTOMER WAS VOMITING, UNABLE TO STAND, DISORIENTED DEHYDRATED AND DIZZY PRIOR TO HOSPITALIZATION. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
180748 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 27 YR Hospitalization