FDA Adverse Event
Injury
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR
MDR report key: 3080105
·
Received April 26, 2013
Report
- Report Number
- 2938836-2013-00988
- Event Type
- Injury
- Date Received
- April 26, 2013
- Date of Event
- February 4, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED THE LEAD WAS RETURNED IN TWO SEGMENTS. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE LEAD WAS OTHERWISE NORMAL.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT WAS HOSPITALIZED DUE TO INAPPROPRIATE SHOCKS. THE IEGM SHOWED NOISE. THE LEAD WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 180747 | DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7122Q/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Required Intervention |