FDA Adverse Event
Injury
Summary report: N
CURRENT RF DR
MDR report key: 3080083
·
Received April 26, 2013
Report
- Report Number
- 2938836-2013-00933
- Event Type
- Injury
- Date Received
- April 26, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- P910023
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
NO MEDWATCH FORM WAS RECEIVED. (B)(6). DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
Additional Manufacturer Narrative · 1
THE REPORTED PREMATURE BATTERY DEPLETION WAS CONFIRMED IN THE LABORATORY. BASED ON THE AVAILABLE PARAMETER AND USAGE INFORMATION, A LONGEVITY CALCULATION WAS PERFORMED AND WAS FOUND TO BE BELOW THE EXPECTED LIMITS. THE DEVICE WAS TESTED ON THE BENCH AND HIGH INPUT CURRENT WAS OBSERVED ON THE HYBRID. THE HIGH INPUT CURRENT WAS CAUSED BY AN ANOMALOUS COMPONENT ON THE HYBRID.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WAS EXPLANTED DUE TO SUSPECTED PREMATURE BATTERY DEPLETION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 180828 | CURRENT RF DR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 2207-36 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |