UNICEL® DXC 600 SYNCHRON® SYSTEM
Report
- Report Number
- 2050012-2013-00294
- Event Type
- Malfunction
- Date Received
- April 26, 2013
- Date of Event
- March 28, 2013
- Report Date
- March 28, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FIELD SERVICE ENGINEER (FSE) VERIFIED SYSTEM PERFORMANCE AND DID NOT OBSERVE ANY SYSTEM ISSUES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. THE REAGENT CARTRIDGE THAT WAS NEAR EMPTY HAD BEEN DISCARDED BY THE CUSTOMER. THE CUSTOMER INDICATED REPLACING THE NEAR EMPTY CREATININE CARTRIDGE RESOLVED THE ISSUE. A DEFINITIVE CAUSE OF THE INCIDENT IS UNKNOWN.
THE AFFILIATE REPORTED THE CUSTOMER ALLEGED ALL THREE LEVELS OF QUALITY CONTROL (QC) FOR CREATININE (CR-S) WERE OUT OF SPECIFICATION INVOLVING THE UNICEL DXC 600 SYNCHRON SYSTEM USED IN CONJUNCTION WITH THE CREATININE REAGENT. THE CUSTOMER STATED THE CREATININE REAGENT CARTRIDGE WAS NEAR EMPTY AND REPLACED THE CARTRIDGE AND PERFORMED CONTROLS; ALL THREE QC LEVELS WERE WITHIN RANGE. THE CUSTOMER THEN ANALYZED ONE PATIENT'S SAMPLE AND OBTAINED A FALSELY LOW CREATININE (CR-S) RESULT OF <27 UMOL/L. THE CUSTOMER REANALYZED THE SAMPLE ON AN ALTERNATE SYSTEM AND RECOVERED A HIGHER RESULT OF 68 UMOL/L. THE CUSTOMER THEN REANALYZED THE SAMPLE ON THE ORIGINAL SYSTEM AND OBTAINED A VALUE OF <27 UMOL/L. THE CUSTOMER RETESTED THE SAMPLE AGAIN, ON THE ALTERNATE SYSTEM, AND OBTAINED A RESULT OF 63 UMOL/L WHICH WAS REPORTED TO THE HOSPITAL. THE CUSTOMER STATED THE ERRONEOUS RESULT WAS NOT RELEASED OUT OF THE LABORATORY. THERE WAS NO PATIENT CONSEQUENCE ASSOCIATED WITH THIS EVENT. SERVICE WAS REQUESTED AND A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 182900 | UNICEL® DXC 600 SYNCHRON® SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR |