MEDRAD VERIS MR VITAL SIGNS MONITOR
Report
- Report Number
- 2520313-2013-00019
- Event Type
- Malfunction
- Date Received
- April 19, 2013
- Date of Event
- March 25, 2013
- Report Date
- March 25, 2013
- Manufacturer
- MEDRAD
- Product Code
- MHX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
BAYER R AND I SERVICE REPLACED THE POWER CHARGER AND POWER SUPPLY CABLE. BAYER R AND I PRODUCT ANALYSIS RECEIVED AND EXAMINED THE DC POWER CABLE AND POWER SUPPLY. THE EXAMINATION OF THE LEMO CONNECTOR AND THE ADJOINING WIRES DETERMINED THAT ONE OF THE WIRES HAD CROSSED OVER AND SHORTED WITH ANOTHER CONDUCTOR TO THE POINT WHERE THE INSULATION OF THE TWO CONDUCTORS BECAME CHARRED. A SUBSEQUENT INVESTIGATION OF SIMILAR COMPLAINTS CONCLUDED THAT WHEN CERTAIN COMBINATIONS OF CONDUCTORS SHORT TOGETHER AT THE CABLE END FARTHEST FROM THE POWER SUPPLY, THIS CAN CREATE A CONDUCTION PATHWAY THAT THE POWER SUPPLY DOES NOT RECOGNIZE AS A SHORT CIRCUIT AND SHUT DOWN. CONSEQUENTLY, THE POWER SUPPLY CONTINUES TO ENERGIZE THE CABLE WHICH CAN LEAD TO LOCALIZED HEATING AND MELTING OF THE POWER CABLE. ON (B)(4) 2013, BAYER HEALTHCARE DISTRIBUTED A FIELD SAFETY NOTICE RECALLING ALL 25-FOOT DC POWER CABLES SHIPPED WITH MEDRAD VERIS MR VITAL SIGNS MONITORS, OR THOSE PROVIDED BY BAYER SERVICE. THESE POWER CABLES ARE BEING RECALLED DUE TO A LATENT DESIGN RELIABILITY ISSUE AND THE POTENTIAL FOR SHORTING WHICH CAN RESULT IN HEATING/MELTING OF THE CABLE JACKET.
THE SITE REPORTED THE FOLLOWING: THE POWER SUPPLY WAS NOT CHARGING THE MONITOR BATTERIES. WHILE ON SITE, A BAYER R AND I SERVICE REP NOTICED THE DC POWER CABLE BECAME VERY HOT TO TOUCH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 168969 | MEDRAD VERIS MR VITAL SIGNS MONITOR | MONITOR, PHYSIOLOGICAL PATIENT | MHX | MEDRAD | 3011996 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |