FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3073716
·
Received April 23, 2013
Report
- Report Number
- 3004209178-2013-92719
- Event Type
- Malfunction
- Date Received
- April 23, 2013
- Date of Event
- March 29, 2013
- Report Date
- April 10, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP ALARM DURING THE PRIME TEST DUE TO LOOSE MOTOR DRIVE SUPPORT DISK.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT THE INSULIN PUMP HAD INSULIN SQUIRTING OUT AND ALARMING DURING MANUAL PRIME. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 173447 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR |