FDA Adverse Event Malfunction Summary report: N

RESERVOIR 1.8ML

MDR report key: 3073696 · Received April 23, 2013

Report

Report Number
3004209178-2013-92718
Event Type
Malfunction
Date Received
April 23, 2013
Date of Event
March 29, 2013
Report Date
March 29, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K001828
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSPECTED ONE OPENED AND USED RESERVOIR. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIR PASSED PER INSPECTION. NO OCCLUDED ANOMALIES WERE FOUND DURING ANALYSIS.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED THAT HIS INSULIN PUMP IS MALFUNCTIONING. CUSTOMER STATED THAT HE RECEIVED A RECERTIFIED REPLACEMENT AND 3 DAYS LATER RECEIVED A MOTOR ERROR ALARM. ASSISTED CUSTOMER WITH MANUAL PRIME AND HE STATE THAT HE CAN SEE INSULIN BUILDING UP RIGHT AROUND THE QR NO DRIPS AND THE INSULIN PUMP DID NOT ALARMED. CUSTOMER ALSO STATED THAT HE SPOKE WITH HIS ATTORNEY AND SENT 2 INFUSION SETS THAT WERE ON THE INSULIN PUMP AT THE TIME OF MOTOR ERROR ALARMED FOR DOCUMENTATION. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
175457 RESERVOIR 1.8ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-326A

Patients

Seq Age Sex Outcome Treatment
1 31 YR UNOMEDICAL INSULIN INFUSION SET