FDA Adverse Event Malfunction Summary report: N

COULTER® LH 750 HEMATOLOGY ANALYZER

MDR report key: 3073675 · Received April 23, 2013

Report

Report Number
1061932-2013-00631
Event Type
Malfunction
Date Received
April 23, 2013
Date of Event
March 29, 2013
Report Date
March 29, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K011342
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE HAD OBSERVED THAT THE SOURCE OF THE LEAK WAS AT THE PROBE WIPE ASSEMBLY. THE FSE REPLACED THE AFFECTED PROBE WIPE ASSEMBLY. NO FURTHER EVIDENCE OF LEAKING WAS OBSERVED. THE CAUSE OF THE LEAK WAS THE PROBE WIPE ASSEMBLY. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO BECKMAN COULTER THAT THE COULTER LH 750 HEMATOLOGY ANALYZER WAS LEAKING. THE VOLUME OF THE LEAK WAS APPROXIMATELY 8 ML AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE ANALYZER DID NOT GENERATE ANY ERROR MESSAGES OR FLAGS AS A RESULT OF THIS EVENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE), CONSISTING OF GLOVES AND A LABORATORY COAT AT THE TIME OF THE OCCURRENCE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. NO ONE CAME IN CONTACT WITH THE FLUID. MEDICAL ATTENTION WAS NOT SOUGHT. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION WITH THIS EVENT. THERE WAS NO DEATH OR INJURY RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
173326 COULTER® LH 750 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER LH 750 N/A