FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3073653 · Received April 23, 2013

Report

Report Number
2032227-2013-01504
Event Type
Injury
Date Received
April 23, 2013
Date of Event
April 3, 2013
Report Date
April 4, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT HE WAS HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS, WITH BLOOD GLUCOSE LEVELS GREATER THAN 600 MG/DL. THE CUSTOMER ALSO EXPERIENCED NAUSEA, VOMITING AND DEHYDRATION. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE INSULIN PUMP PASSED THE PRIME TEST, BUT THERE WAS NO TUBING CLAMP FOR THE HIGH PRESSURE TEST. THE CUSTOMER REQUESTED A REPLACEMENT INSULIN PUMP. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
174375 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAS

Patients

Seq Age Sex Outcome Treatment
1 53 YR Hospitalization