FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK ® COMPACT TEST DRUM

MDR report key: 3073505 · Received April 23, 2013

Report

Report Number
1823260-2013-02482
Event Type
Malfunction
Date Received
April 23, 2013
Date of Event
March 28, 2013
Report Date
May 22, 2013
Manufacturer
ROCHE DIAGNOSTICS
Product Code
LFR
PMA / PMN Number
K031755
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA. DEVICE RECEIVED IN A CONDITION WHICH MADE ANALYSIS IMPOSSIBLE. UNABLE TO TEST BECAUSE AN EMPTY VIAL WAS RETURNED.

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA.

Description of Event or Problem · 1

CALLER REPORTED COMPACT PLUS SYSTEM BLOOD GLUCOSE RESULT OF LO, WHICH ON THE SYSTEM INDICATES A RESULT OF LESS THAN 10 MG/DL. SHE DRANK ORANGE JUICE. SAME SYSTEM RETEST RESULT WAS 240 MG/DL WITHIN 10 MINUTES. NO ADVERSE EVENT WAS REPORTED. A REQUEST WAS MADE FOR THE RETURN OF THE METER AND STRIPS, REPLACEMENT SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
174327 ACCU-CHEK ® COMPACT TEST DRUM BLOOD GLUCOSE MONITORING TEST STRIPS LFR ROCHE DIAGNOSTICS NA 20734543

Patients

Seq Age Sex Outcome Treatment
1 080 YR LASIX| LISINOPRIL| AMLODIPINE| SIMVASTATIN| LANTUS| CLONIDINE| PLAVIX| NOVOLOG| REGLAN| ISOSORB