FLOGARD
Report
- Report Number
- 1416980-2013-10152
- Event Type
- Malfunction
- Date Received
- April 23, 2013
- Date of Event
- April 1, 2013
- Report Date
- April 4, 2013
- Manufacturer
- SHARP CORPORATION OSAKA JAPAN
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A FLO-GARD INFUSION PUMP WITH AN F-49 ALARM WAS CONFIRMED AND REPAIRED BY BAXTER SERVICE PERSONNEL ONSITE AT THE CUSTOMER FACILITY. THE CAUSE OF THE REPORTED CONDITION WAS DETERMINED TO BE BLOWN FUSES WHICH CAUSED THE BATTERY TO GO INTO A STATE OF DEEP DISCHARGE. THE FUSES AND MAIN BATTERY WERE REPLACED TO CORRECT THIS CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
THE FACILITY REPRESENTATIVE REPORTED A FLO-GARD INFUSION PUMP WITH AN F-49 ALARM. IT IS UNKNOWN WHEN, OR IN WHICH CARE AREA, THIS EVENT OCCURRED. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO PATIENT INVOLVEMENT, PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 175069 | FLOGARD | PUMP, INFUSION | FRN | SHARP CORPORATION OSAKA JAPAN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |