FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 3073492 · Received April 23, 2013

Report

Report Number
0001811755-2013-00879
Event Type
Malfunction
Date Received
April 23, 2013
Date of Event
March 29, 2013
Report Date
April 5, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY FOR VISUAL INSPECTION CORROSION WAS FOUND TO BE BUILT UP THROUGHOUT THE INSIDE OF THE HANDPIECE. A BEARING WAS ALSO BROKEN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CORE IMPACTION DRILL HEATED UP DURING USE. THERE WAS NO PATIENT OR USER INJURY REPORTED AND NO ADVERSE CONSEQUENCES ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
175067 CORE IMPACTION DRILL DRIVER, WIRE, AND BONE DRILL, DZJ STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1