FDA Adverse Event Malfunction Summary report: N

LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER

MDR report key: 3073474 · Received April 23, 2013

Report

Report Number
3005075853-2013-01938
Event Type
Malfunction
Date Received
April 23, 2013
Date of Event
April 1, 2013
Report Date
April 2, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED FOR ANALYSIS AND UPON INSPECTION THE JAWS WERE FOUND TO BE IN A YIELDED CONDITION MAKING THE DEVICE NON-FUNCTIONAL. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT. ADDITIONAL INFORMATION REQUESTED AND RECEIVED: THE DEVICE WAS FIRED ON A VESSEL. WAS THE CLIP FULLY ADVANCED INTO THE JAWS PRIOR TO FIRING? YES. WAS THERE ANY TORQUING OR TWISTING OF THE DEVICE PRESENT AT THE TIME OF FIRING? NO. WAS ANY UNEXPECTED RESISTANCE FELT WHILE FIRING THE TRIGGER? NO. WERE ANY UNEXPECTED NOISES HEARD? NO. DID ANYTHING UNEXPECTED HAPPEN PRIOR TO THIS INCIDENT? NO. YES. THE DEVICE WAS FIRED OUT OF THE PATIENT, BUT THE STAPLE WAS STILL MALFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE STAPLES WERE MALFORMED WHEN FIRST AND SECOND FIRING. CHANGED TO ANOTHER ONE TO COMPLETE THE PROCEDURE. NO ADVERSE EVENT ON THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
175042 LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. UNK J4CL3U

Patients

Seq Age Sex Outcome Treatment
1