FDA Adverse Event Injury Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3073390 · Received April 23, 2013

Report

Report Number
2531779-2013-04929
Event Type
Injury
Date Received
April 23, 2013
Report Date
March 27, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2013 THE REPORTER CONTACTED ANIMAS ALLEGING THAT THE PATIENT'S BLOOD GLUCOSE (BG) HAD BEEN "HIGH" (>600MG/DL) FOR THREE DAYS AND THE PATIENT WAS TAKEN TO THE ER ON (B)(6) 2013. THE REPORTER NOTED THAT THE PATIENT HAS CHANGED OUT THE INFUSION SET AND BG COMES DOWN WITH BOLUS BUT THEN GOES UP AGAIN. THE PATIENT'S BG AT THE ER WAS REPORTEDLY 512MG/DL WITH KETONES AND NAUSEA. THE PATIENT WAS REPORTEDLY TREATED WITH BOLUSES VIA THE PUMP AND IV INSULIN DRIP FOR FLUIDS. THE PATIENT WAS TOLD IN THE ER THAT SHE MAY HAVE HAD A VIRUS. THE PATIENT REPORTEDLY REMAINS IN INSULIN PUMP THERAPY. CUSTOMER TECHNICAL SUPPORT REVIEWED THE PUMP WITH THE REPORTED AND FOUND ALL SETTINGS AND HISTORIES ARE CORRECT; THERE WAS NO PUMP DEFECT FOUND DURING TROUBLESHOOTING. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT ALLEGEDLY EXPERIENCED HYPERGLYCEMIA REQUIRING MEDICAL INTERVENTION WHILE USING INSULIN PUMP THERAPY. IT IS UNCLEAR AT THIS TIME OF A POSSIBLE VIRUS WAS THE SOLE CAUSE OF THE REPORTED BG EXCURSIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
174826 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 16 YR Life Threatening| R