FDA Adverse Event
Malfunction
Summary report: N
SECURE II MED/SURG BED
MDR report key: 3073369
·
Received April 23, 2013
Report
- Report Number
- 0001831750-2013-03675
- Event Type
- Malfunction
- Date Received
- April 23, 2013
- Date of Event
- April 2, 2013
- Report Date
- April 2, 2013
- Manufacturer
- STRYKER MEDICAL-KALAMAZOO
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
IT WAS REPORTED VIA REPAIR WORK ORDER THAT THE BED WAS NOT LOWERING FULLY DUE TO BURN IN LIMITS. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 174820 | SECURE II MED/SURG BED | BED, AC-POWERED ADJUSTABLE HOSPITAL | FNL | STRYKER MEDICAL-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |