FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3073258 · Received April 23, 2013

Report

Report Number
2531779-2013-04922
Event Type
Malfunction
Date Received
April 23, 2013
Report Date
March 26, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE DISPLAY SCREEN WAS DIM AND DISCOLORED. A TEST SCREEN WAS PLACED ON THE PUMP AND THE DISPLAY SHOWED NORMALLY. UNRELATED TO THE COMPLAINT, THE PUMP'S HISTORY SHOWED A REBOOT WITH A TIME AND DATE RESET TO FACTORY DEFAULT. DURING TESTING, THE PUMP WENT THROUGH THE REWIND, LOAD, AND PRIME STEPS WITH NO ALARMS. THE PUMP WAS EXERCISED FOR 24 HOURS ON A 1 UNIT PER HOUR BASAL RATE SUCCESSFULLY. THE BATTERY WAS REMOVED FROM THE PUMP FOR SIX HOURS AND THE TIME AND DATE RESET TO FACTORY DEFAULT. THE PUMP WAS OPENED TO INVESTIGATE, AND THE INTERNAL BATTERY WAS FOUND TO BE LEAKING. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A DIM AND DISCOLORED SCREEN. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
175231 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 58 YR