FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3073247 · Received April 23, 2013

Report

Report Number
1416980-2013-10110
Event Type
Malfunction
Date Received
April 23, 2013
Date of Event
April 2, 2013
Report Date
April 2, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A FLO-GARD INFUSION PUMP WITH AN F-6 ALARM WAS CONFIRMED AND REPAIRED BY BAXTER SERVICE PERSONNEL ONSITE AT THE CUSTOMER FACILITY. THE CAUSE OF THE REPORTED CONDITION WAS DETERMINED TO BE A DEFECTIVE MAIN BATTERY. THE MAIN BATTERY WAS REPLACED TO CORRECT THIS CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A FLO-GARD INFUSION PUMP WITH AN F-6 ALARM. IT IS UNKNOWN WHEN, OR IN WHICH CARE AREA, THIS EVENT OCCURRED. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO PATIENT INVOLVEMENT, PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
174287 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1