FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3073075
·
Received April 8, 2013
Report
- Report Number
- 3008642652-2013-00949
- Event Type
- Malfunction
- Date Received
- April 8, 2013
- Date of Event
- March 26, 2013
- Report Date
- April 3, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CABLE) HAS BEEN CONFIRMED. UPON EVAL, THE CABLE CONNECTING THE REAR THERAPY ELECTRODE (TE) AND THE DISTRIBUTION NODE (DN) WAS PULED FROM THE DN STRAIN RELIEF, DAMAGING INTERNAL WIRES. THE ROOT CAUSE OF THE DAMAGED CABLE AND WIRES CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CABLE AND WIRES. THE PT RECEIVED A REPLACEMENT BELT.
Description of Event or Problem · 1
A (B)(6) MALE PT'S WIFE CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT ONE OF THE CABLES ON THE PT'S BELT WAS DAMAGED. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 142821 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR |