FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3073070
·
Received April 8, 2013
Report
- Report Number
- 3008642652-2013-00944
- Event Type
- Malfunction
- Date Received
- April 8, 2013
- Date of Event
- March 1, 2013
- Report Date
- April 3, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK ELECTRODE BELT ALARMS) HAS BEEN CONFIRMED. UPON INVESTIGATION, THE ELECTRODE BELT FAILED AN ECG FALL-OFF TEST. THE CAUSE FOR THE FAILURE WAS ISOLATED TO A BROKEN WHITE (DRVN_GND) WIRE IN THE TRUNK CABLE CONNECTOR. THE ROOT CAUSE FOR THE OPEN WIRE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE ON THE TRUNK CABLE CONNECTOR. NO ADVERSE EVENT RESULTED FROM THE OPEN WIRE. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(6) MALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HE WAS RECEIVING CHECK ELECTRODE BELT ALARMS. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 144431 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |