FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3000 SYSTEM
MDR report key: 3073068
·
Received April 8, 2013
Report
- Report Number
- 3008642652-2013-00929
- Event Type
- Malfunction
- Date Received
- April 8, 2013
- Date of Event
- March 8, 2013
- Report Date
- April 2, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. UPON EVAL, THE GREEN (BELT DISCHARGE) WIRE INSULATION WAS BREAKING DOWN UNDER LOW CURRENT INSIDE THE CABLE CONNECTING ECG ELECTRODES C AND D. THE CAUSE OF THE HI-POT TEST FAILURE IS THE DEFECTIVE WIRE. THE ROOT CAUSE OF THE DEFECTIVE WIRE CANNOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE WIRE. THE LAST PT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.
Description of Event or Problem · 1
REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, BELT SN (B)(4) FAILED INCOMING HI-POT TEST. THE LAST PT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 144321 | LIFEVEST WCD 3000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |