FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3073055 · Received April 8, 2013

Report

Report Number
3008642652-2013-00932
Event Type
Malfunction
Date Received
April 8, 2013
Date of Event
March 11, 2013
Report Date
April 2, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. AS RECEIVED THE MONITOR FAILED INCOMING PULSE TESTING WITH ASSOCIATED CODE 203 - PULSE TEST FAULT AND CODE 102 - CHARGE PROFILE FAULT. THE CAUSE FOR THE CODE 203 AND CODE 201 IS A BROKEN NEGATIVE LEAD ON HIGH-VOLTAGE CAPACITOR C22 ON THE DEFIBRILLATOR BOARD. THE ROOT CAUSE FOR THE BROKEN LEAD ON C22 COULD NOT BE POSITIVELY IDENTIFIED, BUT THE DAMAGE LIKELY RESULTED FROM THE MONITOR IMPACTING A HARD SURFACE. ROOT CAUSE ANALYSIS OF SIMILAR EVENTS FOUND THAT EXCESSIVE STRAIN ON THE PRINTED CIRCUIT ASSEMBLY (PCA) FROM SEVERE IMPACT CAN CAUSE FRACTURED SOLDER CONNECTIONS. (B)(4). NO ADVERSE EVENT RESULTED FROM THE BROKEN LEADS ON C22. THE LAST PATIENT TO USE THIS MONITOR DID NOT REPORT ANY DEFICIENCIES.

Description of Event or Problem · 1

REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, MONITOR SN (B)(4) FAILED INCOMING PULSE TESTING WITH ASSOCIATED CODE 203 AND CODE 102. THE LAST PATIENT TO USE THIS MONITOR DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
144293 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA