FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3073019
·
Received April 8, 2013
Report
- Report Number
- 3008642652-2013-00888
- Event Type
- Malfunction
- Date Received
- April 8, 2013
- Date of Event
- March 12, 2013
- Report Date
- March 28, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CABLE) WAS CONFIRMED. UPON INVESTIGATION THE CABLE THAT CONNECTS THE DISTRIBUTION NODE (DN) TO THE REAR THERAPY ELECTRODES WAS PULLED FROM THE STRAIN RELIEF. THE ROOT CAUSE FOR THE STRAINED CABLE WAS EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED ELECTRODE BELT. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A ZOLL PATIENT SERVICE REPRESENTATIVE (PSR) CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT A (B)(6) MALE PATIENT'S ELECTRODE BELT HAD A DAMAGED CABLE. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 142723 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR |