FDA Adverse Event Malfunction Summary report: N

PROMUS ELEMENT¿ PLUS

MDR report key: 3072913 · Received April 23, 2013

Report

Report Number
2134265-2013-02464
Event Type
Malfunction
Date Received
April 23, 2013
Date of Event
February 12, 2013
Report Date
March 25, 2013
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
P110010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE IS COMBINATION PRODUCT DEVICE EVALUATED BY MFR: A VISUAL AND MICROSCOPIC EXAMINATION IDENTIFIED PROXIMAL STENT DAMAGE. THE FIRST TWO ROWS ON THE PROXIMAL END OF THE STENT WERE RAISED UP. THIS DAMAGE IS CONSISTENT WITH THE STENT MEETING RESISTANCE UPON ADVANCEMENT. THE BALLOON AND THE TIP OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO ANY POSITIVE PRESSURE. NO KINKS OR DAMAGE WERE NOTICED ALONG THE SHAFT OF THE DEVICE. NO OTHER ISSUES WERE NOTED WITH THE DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

THIS COMPLAINT IS NOW REPORTABLE BASED ON THE ANALYSIS COMPLETED ON (B)(4) 2013. IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION THE DEVICE WAS UNABLE TO CROSS THE LESION. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT CIRCUMFLEX CORONARY ARTERY (LCX). THE LESION WAS PREDILATED AND THEN A 2.50X24MM PROMUS ELEMENT STENT DELIVERY SYSTEM (SDS) WAS ADVANCED TO THE LCX; HOWEVER, THE DEVICE WAS UNABLE TO CROSS THE LESION. THE DEVICE WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S CURRENT CONDITION IS STABLE. HOWEVER, RETURNED PRODUCT ANALYSIS REVEALED STENT DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
175093 PROMUS ELEMENT¿ PLUS STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC - GALWAY H7493918424250 15303271

Patients

Seq Age Sex Outcome Treatment
1 76 YR